The NanoCool™ Injector is a single syringe infusion/withdrawal programmable syringe pump with a flow rate range of 367.4 pl/min to 1.906 ml/min with 11 lbs of adjustable force across the entire flow rate range. It can accommodate syringes from 50 µl to 500 µl.
When injecting stem cells viability of cells can be a major issue. The injection system and sample vortex holder are held at 18°C. The cell delivery is totally integrated in it operation through the NanoCool's microprocessor system: power, temperature, injection volumes and flow ramping are all controlled by the NanoCool's program. Keeping the cells at 18°C instead of 37°C reduces O2consumption by an average of 76%.
The NanoCool™ Injector is very easy to use with an LCD color touch screen and icon interface. This pump allows you to create, save and run simple to complex Methods without a PC. The Method profiles available on this pump are Constant Rate, Ramp, Stepped, Bolus, Concentration, Gradient (binary or ternary), Pulse and Autofill. Delays can be set using the real and relative time clocks.
The NanoCool™ Injector is a single syringe infusion/withdrawal programmable syringe pump with a flow rate range of 367.4 pl/min to 1.906 ml/min with 11 lbs of adjustable force across the entire flow rate range. It can accomodate syringes from 50 µl to 500 µl.
Features:
Applications:
When injecting stem cells viability of cells can be a major issue. The injection system and sample vortex holder are held at 18°C. The cell delivery is totally integrated in it operation through the NanoCool's microprocessor system: power, temperature, injection volumes and flow ramping are all controlled by the NanoCool's program. Keeping the cells at 18°C instead of 37°C reduces O2 consumption by an average of 76%.
The NanoCool™ Injector is very easy to use with an LCD color touch screen and icon interface. This pump allows you to create, save and run simple to complex Methods without a PC. The Method profiles available on this pump are Constant Rate, Ramp, Stepped, Bolus, Concentration, Gradient (binary or ternary), Pulse and Autofill. Delays can be set using the real and relative time clocks.
The Message Area of the touch screen is used to display helpful instructions for the currently displayed screen. It is also used to display error or warning messages to indicate problem conditions in a Method or error conditions during pump operation. The Run Screen shows all of the pump parameters on one screen for easy review.
The software is organized into three main navigational branches, the quick start Methods, user-defined Methods, and system settings. You can control operations directly through the touch screen or remotely from an independent computer or device via the external I/O interface.
The NanoCool™ Injector has a footswitch input, RS-232 and USB serial ports for computer control, RS-485 ports for daisy chaining pumps and Digital I/O for external control via an independent computer or device. There is also an option for daisy chaining pumps through the RS-232 (RJ-11) ports. This option must be ordered at the time the pump is ordered.
This pump consists of a control unit, an injection unit, a 6 foot cable to connect the two units and a footswitch
Specifications | 88-1050 |
---|---|
Accuracy | ±0.5% |
Classification | Class I |
Dimensions Control Box LxDxH in cm | 12 x 8.5 x 4.38 (30.5 x 21.6 x 11.1) |
Dimensions Mechanism LxDxH in cm | 2.5 x 2.0 x 7.5 (6.35 x 5.08 x 19.05) |
Display | 4.3" WQVGA TFT Color Display with Touchpad |
Drive Motor | 1.8° Stepper Motor |
Environmental Humidity | 20% to 80% RH, non condensing |
Environmental Operating Temperature English | 40°F to 104°F |
Environmental Operating Temperature Metric | 4°C to 40°C |
Environmental Storage Temperatue English | 14°F to 158°F |
Environmental Storage Temperatue Metric | -10°C to 70°C |
Flow Rate Maximum | 367.4 pl/min using 50 µl syringe |
Flow Rate Minimum | 1.906 ml/min using 500 µl syringe |
I O TTL Connectors | 15 pin D-Sub Connector |
Input Power | 75 W, 0.5 A fuse |
Installation Category | II |
Max Linear Force | 11 lbs @ 100% Force Selection |
Mode of Operation | Continuous |
Motor Drive Control | Microprocessor with 1/16 microstepping |
Net Weight Control Box English | 4.55 lb |
Net Weight Control Box Metric | 2.06 kg |
Net Weight Remote Mechanism English | 1.01 lb |
Net Weight Remote Mechanism Metric | 0.458 kg |
No of Syringes | 1 |
Non Volatile Memory | Storage of all settings |
Number of Microsteps per one rev of Lead Screw | 3,200 |
Pollution Degree | 1 |
Pump Configuration | Remote |
Pump Function | Infuse/Withdraw, Programmable |
Pusher Travel Rate Maximum | 228.97 mm/min |
Pusher Travel Rate Minimum | 0.433 µm/min |
RS 232 Connectors | 9 pin D-Sub Connector |
Regulatory Certifications | CE, ETL (UL, CSA), WEE, EU RoHS & CB Scheme |
Step Rate Maximum | 52 µsec/µstep |
Step Rate Minimum | 27.5 sec/µstep |
Syringe Rack Type | Standard Rack |
Syringe Size Maximum | 500 µl |
Syringe Size Minimum | 50 µl |
USB Connectors | Type B |
Voltage Range | 100-240 VAC, 50/60 Hz |